The list below contains Medical Device’s Field Corrective Action for the month of August 2025. This list provides field corrective action reporting information to MDA by establishment in accordance to requirement stipulated in Section 41, Act 737 and Regulation 6 of Medical Device (Duties and Obligations of Establishment) Regulations 2019.

Field Corrective Action (FCA) is part of a post-marketing risk assessment measure to ensure the continued safe use of medical devices in the Malaysia market. It is a corrective or preventive action in relation to a medical device which may include –

  • The return of the medical device to the establishment;
  • Modification of the medical device;
  • Exchange of the medical device;
  • Destruction of medical device; or
  • Specific advice on the use of the medical device. 

If you have experienced any safety and performance issue with one of the medical devices listed for FCA action, please seek advice from your medical device authorised representative as soon as possible.

  MEDICAL DEVICE FIELD CORRECTIVE ACTION LISTING AUGUST 2025

No.

Date Received

Title of FCA

Affected Medical Device

MDA Reference Number

MDA Registration Number

Local Establishment Contact Detail

1.

21/08/2025

Catheter Foley 2 Way Balloon with potential for water in the catheter balloon not being able to be removed and catheter balloon not being able to deflate

CATHETER FOLEY 2 WAY BALLOON

MDA/FCA/PX1229-80546460-2025

GB36726735318

MYMEDIC INNOVATION SDN BHD

[email protected]

2.

14/08/2025

Datascope Cardiosave Hybrid and Rescue Intra-Aortic Balloon Pumps (IABP) System Over Temperature Field Action

CARDIOSAVE INTRA-AORTIC BALLOON PUMP

 MDA/FCA/P1365-48897212-2025

GC2992923-137663

RBD HEALTHCARE SDN BHD

[email protected]

3.

07/08/2025

036 FSCA ACHC 24-07 2nd Follow Up Atellica CH Revised C Reactive Protein (RCRP) | Incorrect Software Flagging for the Atellica CH Revised C Reactive Protein (RCRP) Assay

ATELLICA® CH REVISED C‑REACTIVE PROTEIN (RCRP)

 MDA/FCA/P1371-79785116-2025

IVDB10942623-142160

SIEMENS HEALTHCARE SDN. BHD.

[email protected]

4.

07/08/2025

037 CN POC 25-015 Chek Stix Liquid QC Kit | Foreign Material Identified in Quality Control Product

AUTOMATED URINE CHEMISTRY ANALYZER

 MDA/FCA/P1372-62421682-2025

IVDB36827210418

SIEMENS HEALTHCARE SDN. BHD. [email protected]

5.

08/08/2025

Field Safety Notice Urgent Medical Device Correction – Potential for Loose Universal Surgical          Manipulators on da Vinci X and Xi Systems (ISIFA2022-14-C)

DA VINCI XI SURGICAL SYSTEM, ENDOSCOPIC INSTRUMENT CONTROL SYSTEM, MODEL IS4000

 MDA/FCA/P1373-37231744-2025

GC3810021-64842

DTG MEDICAL SDN BHD

[email protected]

6.

11/08/2025

VITROS® Chemistry Products Ca Slides, Generations 67 and Above, May Experience Increased Incidence of Calibration Failures

VITROS CHEMISTRY PRODUCTS ELECTROLYTES REAGENTS

 MDA/FCA/P1374-48492421-2025

IVDB7599223-142604

T T MEDICAL MANAGEMENT SDN BHD

[email protected]

7.

13/08/2025

MR040/25/S (CSAN), MR042/25/S - Inspection of Magnet Venting System

MAGNETOM SKYRA

 MDA/FCA/P1377-24628432-2025

GB44064488717

SIEMENS HEALTHCARE SDN. BHD.

[email protected]

8.

13/08/2025

MR040/25/S (CSAN), MR042/25/S - Inspection of Magnet Venting System

MAGNETOM PRISMA

  MDA/FCA/P1378-43240356-2025

GB75698797118

SIEMENS HEALTHCARE SDN. BHD.

[email protected]

9.

13/08/2025

MR040/25/S (CSAN), MR042/25/S - Inspection of Magnet Venting System

MAGNETOM VIDA

 MDA/FCA/P1378-65762248-2025

GB75575532218

SIEMENS HEALTHCARE SDN. BHD.

[email protected]

10.

13/08/2025

MR040/25/S (CSAN), MR042/25/S - Inspection of Magnet Venting System

MAGNETOM LUMINA

 MDA/FCA/P1378-91610863-2025

GB3902419-30683

SIEMENS HEALTHCARE SDN. BHD.

[email protected]

11.

13/08/2025

FA-25056- on the DxC 500i Clinical Analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC constituent tests, the sample status of the CC test will remain “In Progress” with no error events.

DxC 500i AU Chemistry Module

 MDA/FCA/P1382-16528305-2025

IVDA8913224-171351

BECKMAN COULTER MALAYSIA SDN. BHD.

[email protected]

12.

14/08/2025

MR041/25/S (CSAN), MR042/25/S - Inspection of Magnet Venting System

MAGNETOM SPECTRA

 MDA/FCA/P1383-25259554-2025

GB64827212817

SIEMENS HEALTHCARE SDN. BHD

[email protected]

13.

14/08/2025

038 FSCA ACHC 25-07 Atellica CH Urinary Cerebrospinal Fluid Protein (UCFP) | Falsely Depressed Results for the Atellica CH UCFP Assay. May Act as a Heat Source with Potential to Produce Smoke and/or Fire.

ATELLICA® CH URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP)

 MDA/FCA/P1384-50133025-2025

IVDB5927720-49887

SIEMENS HEALTHCARE SDN. BHD.

[email protected]

14.

18/08/2025

The two housings of Various Clear-Therm Mini HMEF devices separate when a force is applied during handling

BREATHING FILTERS, HMES AND HMEFS

MDA/FCA/P1386-20871349-2025

GB890231283619

INTERSURGICAL SDN BHD

[email protected]

15.

18/08/2025

Updating product-specific Instructions for Use (IFU) for Skull Traction Tong Large

AESCULAP SURGICAL INSTRUMENT NEUROSURGERY

 MDA/FCA/P1387-98630904-2025

GD17535444617

B. BRAUN MEDICAL INDUSTRIES SDN. BHD.

[email protected]

16.

18/08/2025

Urgent Medical Device Correction for the PrisMax Systems and TherMax Blood Warmer Units on PrisMax System alarm T2309, PrisMax System alarm T0830 and Thermax System may be unable to detect the presence of a bag on the Thermax Blood Warmer Unit leading to difficulty setting up therapy, or PrisMax System alarm T2284

PRISMAX

 MDA/FCA/P1388-32386631-2025

GC9127823-124035

VANTIVE SDN BHD

[email protected]

17.

20/08/2025

NOTIFICATION ON POTENTIAL WEAK IMMUNOHISTOCHEMISTRY STAINING WITH FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use (Dako Omnis), GA604

DAKO OMNIS (ATSI)-LYMPHATIC TISSUE AND BONE MARROW

 MDA/FCA/P1390-13117620-2025

IVDC7441223-151113

BITA LIFESCIENCE SDN. BHD.

[email protected]

18.

25/08/2025

FA-AM-FEB2025-305: Alinity M Software Update 1.9.0

ALINITY M SYSTEM

 MDA/FCA/P1392-94280894-2025

IVDB7120020-41682

ABBOTT LABORATORIES (MALAYSIA) SDN. BHD.

[email protected]

19.

25/08/2025

Urgent on replace of HU 35 Tank Fittings

HEATER UNIT HU35

 MDA/FCA/P1393-95249345-2025

GC6292720-45160

RBD HEALTHCARE SDN BHD

[email protected]

20.

26/08/2025

039 FSCA ACHC 25-07 Follow Up Atellica CH Urinary Cerebrospinal Fluid Protein (UCFP) | Potential for Falsely Depressed Results with the Atellica CH UCFP Assay

ATELLICA® CH URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP)

 MDA/FCA/P1395-19923798-2025

IVDB5927720-49887

SIEMENS HEALTHCARE SDN. BHD.

[email protected]

* The information contained in the Medical Device Authority Field Corrective Action database is released under Regulation 7(8) and Regulation 8 (5) of Malaysia’s Medical Device Regulations 2019.