The list below contains Medical Device’s Field Corrective Action for the month of October 2025. This list provides field corrective action reporting information to MDA by establishment in accordance to requirement stipulated in Section 41, Act 737 and Regulation 6 of Medical Device (Duties and Obligations of Establishment) Regulations 2019.

Field Corrective Action (FCA) is part of a post-marketing risk assessment measure to ensure the continued safe use of medical devices in the Malaysia market. It is a corrective or preventive action in relation to a medical device which may include –

  • The return of the medical device to the establishment;
  • Modification of the medical device;
  • Exchange of the medical device;
  • Destruction of medical device; or
  • Specific advice on the use of the medical device. 

If you have experienced any safety and performance issue with one of the medical devices listed for FCA action, please seek advice from your medical device authorised representative as soon as possible.

    MEDICAL DEVICE FIELD CORRECTIVE ACTION LISTING OCTOBER 2025

No.

Date Received

Title of FCA

Affected Medical Device

MDA Reference Number

MDA Registration Number

Local Establishment Contact Detail

1.

01/10/2025

LOGIQ P9, P10 R4.5.7 Enhancement of UGAP Measurement Result LOGIQ P SERIES (LOGIQ P8, LOGIQ P9, LOGIQ P10) (GB4889221-73580)

LOGIQ P SERIES (LOGIQ P8, LOGIQ P9, LOGIQ P10)

 MDA/FCA/P1424-54608372-2025

GB4889221-73580

GE HEALTHCARE SDN. BHD

[email protected]

2.

01/10/2025

LOGIQ P9, P10 R4.5.7 Enhancement of UGAP Measurement Result LOGIQ P SERIES (GB6568122-103525)

LOGIQ P SERIES

 MDA/FCA/P1425-26916987-2025

GB6568122-103525

GE HEALTHCARE SDN. BHD

[email protected]

3.

07/10/2025

FA-25071- Beckman Coulter has confirmed a software anomaly on DxI 9000 analyzers that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number".

ACCESS PROGESTERONE II

 MDA/FCA/P1427-80767855-2025

IVDB9522424-187370

BECKMAN COULTER MALAYSIA SDN. BHD.

[email protected]

4.

07/10/2025

FA-25071- Beckman Coulter has confirmed a software anomaly on DxC 500i analyzers that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number".

DxC 500i AU Chemistry Module

 MDA/FCA/P1428-64045206-2025

IVDA8913224-171351

BECKMAN COULTER MALAYSIA SDN. BHD.

[email protected]

5.

03/10/2025

2024-IGT-PUN-004 Zenition 50 Potential loss of imaging functionality and/or poor image quality

ZENITION 50

 MDA/FCA/P1429-94718851-2025

GC2933619-31252

PHILIPS MALAYSIA SDN BERHAD

[email protected]

6.

07/10/2025

IDS-25-5176- EXP Product Cybersecurity Matter

BD PHOENIX™ AUTOMATED MICROBIOLOGY SYSTEM

 MDA/FCA/P1433-74532171-2025

IVDB10660623-151263

BECTON DICKINSON SDN BHD

[email protected]

7.

14/10/2025

Olympus to Reinforce the IFU

SHOCKPULSE-SE LITHOTRIPSY SYSTEM

 MDA/FCA/P1434-67079874-2025

GC92154340917

OLYMPUS (MALAYSIA) SDN. BHD.

[email protected]

8.

08/10/2025

FA-24068 - Supplier manufacturing nonconformance in PnP collet covers (C20585) from a Dec 2023-Jan 2024 batch causing overly thick flags to rub and damage Z-axis sensors (C20586).

UniCel DxI Access Immunoassay Systems

 MDA/FCA/P1435-79338984-2025

IVDA9233125-201119

BECKMAN COULTER MALAYSIA SDN. BHD.

[email protected]

9.

09/10/2025

Potential mycobacterial

contamination on Heater Unit HU35

HEATER UNIT HU35

 MDA/FCA/P1436-94754693-2025

GC6292720-45160

RBD HEALTHCARE SDN BHD

[email protected]

10.

13/10/2025

Olympus to Provide Online Training Materials on Reprocessing for TJF Duodenoscopes

EVIS EXERA III DUODENOVIDEOSCOPE TJF-Q190V

 MDA/FCA/P1437-46822570-2025

GB4632321-70352

OLYMPUS (MALAYSIA) SDN. BHD.

[email protected]

11.

15/10/2025

CentriMag Pump OOS not fitting in the CentriMag Motor

CENTRIMAG AND PEDIVAS EXTRACORPOREAL BLOOD PUMPING SYSTEM

 MDA/FCA/P1438-83908883-2025

GC37805334117

ABBOTT MEDICAL (MALAYSIA) SDN. BHD.

[email protected]

12.

15/10/2025

HeartMate Controller Backup Battery Fault

HEARTMATE 3 LEFT VENTRICULAR ASSIST SYSTEM

 MDA/FCA/P1439-69051125-2025

GD4151721-74880

ABBOTT MEDICAL (MALAYSIA) SDN. BHD.

[email protected]

13.

22/10/2025

AccuScreen DPOAE Probe - inducing technical distortions on high sound levels

MADSEN ACCUSCREEN

  MDA/FCA/P1442-90983973-2025

GB3401824-162662

ACCESS 2 HEALTHCARE SDN BHD

[email protected]

14.

22/10/2025

Automated Impella Controller (AIC) Cybersecurity Notification

IMPELLA CP® WITH SMARTASSIST® SET

 MDA/FCA/P1442-88938875-2025

GD9474523-140294

DCH AURIGA (M) SDN BHD

[email protected]

15.

17/10/2025

PCS/MS – Potential FLEX-3S2P battery functional failures PORTRAIT VSM VITAL SIGNS MONITOR (GC7676824-159439)

PORTRAIT VSM VITAL SIGNS MONITOR

 MDA/FCA/P1444-42806470-2025

GC7676824-159439

GE HEALTHCARE SDN. BHD

[email protected]

16.

17/10/2025

PCS/MS – Potential FLEX-3S2P battery functional failures CARESCAPE ONE (GC5319024-173964)

CARESCAPE ONE

 MDA/FCA/P1445-24236001-2025

GC5319024-173964

GE HEALTHCARE SDN. BHD

[email protected]

17.

17/10/2025

PCS/MS – Potential FLEX-3S2P battery functional failures B105P/B105M/B125P/B125M/B155M PATIENT MONITOR ( GC5462723-144051)

B105P/B105M/B125P/B125M/B155M PATIENT MONITOR

 MDA/FCA/P1446-71669218-2025

GC3218021-63208

GE HEALTHCARE SDN. BHD

[email protected]

18.

17/10/2025

PCS/MS – Potential FLEX-3S2P battery functional failures CARESCAPE MONITOR B450 (GC52869540718)

CARESCAPE MONITOR B450

 MDA/FCA/P1447-66979970-2025

GC52869540718

GE HEALTHCARE SDN. BHD

[email protected]

19.

17/10/2025

PCS/MS – Potential FLEX-3S2P battery functional failures CARESCAPE B850 B650 B450 (GC7153322-104146

CARESCAPE B850 B650 B450

 MDA/FCA/P1448-54127603-2025

GC7153322-104146

GE HEALTHCARE SDN. BHD

[email protected]

20.

21/10/2025

FA-25054 Customer complaints were received for multiple Bicarbonate reagent lots associated with false high patient results for Bicarbonate. The high results are due to elevated levels of LDH in the patient samples tested. The current Bicarbonate product IFU’s do not advise of LDH interference with this assay.

BECKMAN COULTER ELECTROLYTE REAGENTS

 MDA/FCA/P1450-60137067-2025

IVDB2268977017

BECKMAN COULTER MALAYSIA SDN. BHD.

[email protected]

21.

22/10/2025

Liquid leaking from the “Valve _Assy NO.339” built in CN

Sysmex_Automated Coagulation Analyzer_CN Series

MDA/FCA/P1451-18248256-2025

IVDC6362721-66356

SYSMEX (MALAYSIA) SDN. BHD

[email protected]

22.

22/10/2025

FA1524 CareLink Clinic Dashboard Display Correction

GUARDIAN CONNECT SYSTEM AND ACCESSORIES

 MDA/FCA/P1453-95051837-2025

GB8317423-114689

MEDTRONIC MALAYSIA SDN BHD

[email protected]

23.

23/10/2025

Field Safety Notice on accuracy of Triglyceride, Total Cholesterol and Low-Density Lipoprotein Cholesterol analytes when using the Afinion™ Lipid Panel

AFINION LIPID PANEL

 MDA/FCA/P1454-11351166-2025

IVDB74056332218

ABBOTT DIAGNOSTICS HEALTH SDN BHD

[email protected]

24.

23/10/2025

003 FSCA AIMC 25-06 Follow Up Atellica IM ADVIA Centaur Homocysteine Calibrator | Atellica IM and ADVIA Centaur Homocysteine Calibrator Instability

ATELLICA® IM HOMOCYSTEINE (HCY)

 MDA/FCA/P1455-79820434-2025

IVDB59216279518

SIEMENS HEALTHCARE SDN. BHD.

[email protected]

25.

27/10/2025

2025-PD-CTAMI-107 - Philips CT Systems - Potential for System Shutdown and Expelled Parts

INCISIVE CT

 MDA/FCA/P1462-60923675-2025

GC74519972618

PHILIPS MALAYSIA SDN BERHAD

[email protected]

26.

28/10/2025

Field action for InteliQ Immunology Control to revise the in-use stability claim for anti-cyclic Citrullinated Peptide (anti-CCP).

INTELIQ IMMUNOLOGY CONTROL

 MDA/FCA/P1409-71638251-2025

IVDB9987521-63532

CHEMOPHARM SDN BHD

[email protected]

27.

29/10/2025

EA 9180 Na discrepant results

REFERENCE ELECTRODE

 MDA/FCA/P1463-17401355-2025

IVDB7921024-186374

ROCHE DIAGNOSTICS (M) SDN. BHD

[email protected]

28.

29/10/2025

A problem related to the device failing to start mechanical CPR after the software update to version 1.1.3 occurred

EASY PULSE - MECHANICAL CPR

 MDA/FCA/P1467-69692714-2025

GC8466522-101961

SCHILLER ASIA PACIFIC SDN. BHD

[email protected]

29.

29/10/2025

Hoses of the Catheter Mounts may tear during use and lead to leakages (ErgoStars with smoothbore hoses).

DRAEGER SAVINA FAMILY

 MDA/FCA/P1468-48690636-2025

GC84895244017

DRAEGER MALAYSIA SDN BHD

[email protected]

30.

30/10/2025

Clear-Therm Mini Paediatric HMEF housings separate when a force is applied

BREATHING FILTERS, HMES AND HMEFS

 MDA/FCA/P1469-95019061-2025

GB890231283619

INTERSURGICAL SDN BHD

[email protected]

31.

30/10/2025

FA1511 Rev A_Potential for Suspension of Dynamic Sensing Algorithm Aurora EV-ICD (DVEA3E4)

AURORA EV-ICD™ SYSTEM

 MDA/FCA/P1470-78365963-2025

GD2666524-159481

MEDTRONIC MALAYSIA SDN BHD

[email protected]

* The information contained in the Medical Device Authority Field Corrective Action database is released under Regulation 7(8) and Regulation 8 (5) of Malaysia’s Medical Device Regulations 2019.